Key result
Hemoglobin Cheverly causes unexpectedly low SpO2 that does not predict clinical hypoxemia.
Why the study?
Pulse-oximeter monitoring of oxygenation status may be inaccurate in patients with unstable hemoglobins such as hemoglobin Cheverly, leading to unexpectedly low oxygen saturation readings.
Case Report (n=1)
Pulse-oximeter monitoring of oxygenation status may not be appropriate for patients with unstable hemoglobins like hemoglobin Cheverly, as it can yield falsely low readings.
Answers are grounded in this paper and cite their sources.
Pulse oximetry may mislead in unstable hemoglobins; this case leaves open broader applicability and monitoring alternatives.
Unexpectedly low hemoglobin oxygen saturation as determined by pulse-oximeter analysis was observed in a patient who underwent an elective surgical procedure. Specific hemoglobin derivatives such as carboxyhemoglobin, methemoglobin, and reduced hemoglobin that have been described to lower pulse-oximetry determination of oxygenation were not detected. Absorbance spectra revealed the patient's hemoglobin to be different than that obtained from two normal volunteers. High-pressure liquid chromatographic analysis of the hemoglobin showed an unknown band that comprised 15% of the patient's hemoglobin. DNA sequence analysis showed a point mutation in the second nucleotide of the 45th codon of the beta-globin chain. This mutation encodes for an abnormal beta-chain (beta-45 Phe-->Ser) that has been described as hemoglobin Cheverly. Hemoglobin Cheverly is an unstable hemoglobin that has a similar mutation as the beta-42 Phe-->Ser mutation seen in hemoglobin Hammersmith. Hemoglobin Hammersmith and another unstable hemoglobin, hemoglobin Köln, have previously been described to have unexpectedly low pulse-oximeter-determined oxyhemoglobin levels. That we find hemoglobin Cheverly to result in a similar phenomenon suggests that pulse-oximeter monitoring of oxygenation status may not be appropriate for the unstable hemoglobins. Low pulse-oximeter oxygenation determinations for these hemoglobins do not appear to predict clinical hypoxemia.
No takes yet. Share an insight, caveat, or question.
Hohl et al. (1998) conducted a case report in Hemoglobin Cheverly (n=1). Hemoglobin Cheverly vs. Normal hemoglobin was evaluated on Pulse-oximeter-measured hemoglobin oxygen saturation. Hemoglobin Cheverly, an unstable hemoglobin variant, causes unexpectedly low pulse-oximeter-measured oxygen saturation that does not predict clinical hypoxemia.
Synapse has enriched 10 related papers. Consider them for comparative context: