This is the authors' abstract. We don't add key points for this paper.
Bio-instructive materials that prevent bacterial biofilm formation and drive an appropriate host immune response have the potential to significantly reduce the burden of medical device-associated infections. Since bacterial surface attachment is known to be sensitive to surface topography, we experimentally survey 2,176 combinatorially generated shapes using an unbiased high-throughput micro topographical screen on polystyrene. This identifies topographies that reduce colonization in vitro by up to 15-fold compared with a flat surface for both motile and non-motile bacterial pathogens. Equivalent reductions are achieved on polyurethane, a polymer commonly used in medical devices. Using machine learning methods, a set of design rules based on generalisable descriptors is established for predicting bacteria-resistant micro topographies. In a murine foreign body infection model, anti-attachment topographies are shown to be refractory to P. aeruginosa and to recruit a productive host response, highlighting the potential of simple topographical patterning of non-eluting implants for preventing medical device associated infections.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Romero et al. (2020) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: