Answers are grounded in this paper and cite their sources.
Foil thickness measurements from convergent-beam electron diffraction (CBED) patterns may be easily made using a technique based on the two-beam approximation to dynamical diffraction theory. A previous theoretical study showed that, under certain conditions, multiple-beam interactions could seriously affect the accuracy of the technique by causing systematic errors. In the present study, the precision of a CBED technique for the measurement of foil thickness was studied experimentally by analysing diffraction patterns taken at several locations using four different diffraction vectors in a Co-based alloy. A precision of ∼3% was achieved when measurements from a single location taken under different conditions were compared. For the individual thickness measurements, using patterns from 220 or 311 reflections, the width of the 95% confidence intervals was in no instance greater than 4% of the measured thickness. Values of ξg, which are also obtained from the CBED analysis, showed a 95% confidence interval which was 5–11%, of ξg. Systematic errors associated with the CBED technique arising from multiple-beam interactions were experimentally verified. In the case of the Co-based alloy used in the present study, these errors were large for a 111 reflection. Comparison of CBED thickness measurements with measurements based on the projected width of stacking faults show consistent results, although the estimated error in the fault width determinations is considerably larger than for the CBED results.
No takes yet. Share an insight, caveat, or question.
Allen et al. (1982) studied this question.