Key result
Contemporary clinical software of major vendors demonstrated a mean inter-vendor difference in global longitudinal strain below 1 strain% and a test-retest variability of 5.2–6.3%.
Answers are grounded in this paper and cite their sources.
Contemporary clinical software from major vendors shows marked convergence in GLS measurements, allowing reliable detection of guideline-recommended relative changes when using the same software.
We thank Dr. Kaya for his interest in our recently published study on inter-vendor differences and test-retest reproducibility of endocardial and mid-/full-wall left ventricular global longitudinal strain (GLS).1 Our study demonstrated a marked convergence of GLS measurements over the past years, at least when measurements are obtained with contemporary clinical software of major vendors.1 The mean inter-vendor difference was now clearly below 1 strain% while test-re-test variability remained low and showed no relevant difference among the same vendors (5.2–6.3%).1 Based on these data, we could report a minimal detectable change (MDC, 95% confidence) between 2.0 strain% and 2.9 strain% for repeated measurements with the same vendor and between 3.0 strain% and 4.6 strain% when using any combination of vendors with contemporary clinical software.1 We share the notion of our colleagues that longitudinal follow up of patients requires particularly consistent measurements. Our data prove that guideline recommended relative changes in GLS of 10–15% (equal to 2 to 3 absolute strain % change when departing from a normal strain of 20%) can be reliably detected when using the same software, and that combinations of vendors with contemporary software are already on the edge of reaching this target. Nevertheless, it should be noted, that GLS remains in any way among the most reproducible parameters in echocardiography.
No takes yet. Share an insight, caveat, or question.
Balinisteanu et al. (2026) conducted a letter in Left ventricular global longitudinal strain (GLS) measurement. 2D-speckle tracking global longitudinal strain measurements using contemporary clinical software vs. Different vendors or repeated measurements was evaluated on Inter-vendor differences and test-retest reproducibility of left ventricular global longitudinal strain. Contemporary clinical software of major vendors demonstrated a mean inter-vendor difference in global longitudinal strain below 1 strain% and a test-retest variability of 5.2–6.3%.
Synapse has enriched 10 related papers. Consider them for comparative context: