Key result
Validated LC-MS/MS method reliably quantifies eniluracil and 5-fluorouracil in human plasma.
Why the study?
A reliable and validated method for simultaneous quantification of eniluracil and 5-fluorouracil in human plasma was needed for clinical studies.
A sensitive LC-MS/MS method was developed and validated for the simultaneous quantification of the chemotherapeutic agent 5-fluorouracil and its modulator eniluracil in human plasma.
May enable pharmacokinetic monitoring in 5-FU regimens; leaves open clinical adoption pending prospective validation.
BACKGROUND: 5-fluorouracil (5-FU) has been one of the most widely used chemotherapeutic agents to treat various tumors, and eniluracil (5-ethynyluracil or EU) is being developed as a novel modulator of 5-FU. RESULTS: A simple and sensitive LC-MS/MS method was developed for reliably quantifying both EU and 5-FU in human plasma. The method was validated for EU over a dynamic concentration range from 4.13 ng/ml (LOQ) to 1030 ng/ml and for 5-FU over a dynamic concentration range from 8.61 ng/ml (LOQ) to 1080 ng/ml. The analog, 5-bromouracil, was used as the internal standard for calibration curves and quantitation. Method validation has covered the scope of precision, accuracy, specificity, LOQ, linearity/range, freeze-thaw cycles, benchtop integrity/stability, storage stability, matrix effect, recoveries and so on, in accordance with US FDA bioanalytical method validation guidelines. CONCLUSIONS: The validated method has shown good applicability for clinical studies and may be used for other clinical trials that involve measuring the concentration of EU and 5-FU simultaneously in human plasma and potentially in other similar biological matrices.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2010) studied this question. LC-MS/MS method for quantifying eniluracil and 5-fluorouracil was evaluated on Method validation (precision, accuracy, specificity, LOQ, linearity, stability). A simple and sensitive LC-MS/MS method was successfully developed and validated for reliably quantifying both eniluracil (4.13-1030 ng/ml) and 5-fluorouracil (8.61-1080 ng/ml) in human plasma.
Synapse has enriched one closely related paper. Consider it for comparative context: