Key result
MGuard Stent linked to ~10-fold higher TLR risk vs BMS in STEMI patients.
Why the study?
There are no real world data comparing the MGuard Stent with the bare metal stent in STEMI patients.
Does the MGuard Stent reduce adverse cardiac events compared to bare metal stents in patients with STEMI?
Cohort (n=262)
No
Does the MGuard Stent reduce adverse cardiac events compared to bare metal stents in patients with STEMI?
In real-world STEMI patients, the MGuard Stent is associated with a significantly higher rate of target lesion revascularization compared to bare metal stents, without differences in mortality.
Answers are grounded in this paper and cite their sources.
Single-center retrospective data on MGS versus BMS in STEMI are hypothesis-generating; randomized trials needed before clinical adoption.
AIM: The MGuard Stent (MGS) was designed to prevent distal embolization of thrombus and has been shown to improve microcirculation in ST-elevation myocardial infarction (STEMI). However, there are no real world data comparing it with the bare metal stent (BMS). The aim of this study is to determine the efficacy and safety of the MGS in STEMI in the real world compared to the BMS. METHODS AND RESULTS: In total, 262 patients were included from a single centre, of which 35.9% had an MGS implanted. Two groups of 79 patients were established after propensity score matching, and they were similar in terms of baseline and periprocedural variables. The mean follow-up was 321 ± 12.94 days. There was no difference in mortality (7.6% in both groups), major adverse cardiac events (20.3% vs. 12.7%, P = 0.198), non-cardiac mortality, or non-fatal myocardial infarction (6.3% in both groups). Target lesion revascularization (TLR) was significantly higher in the MGS group (11.4% (9) vs. 1.3% (1) P < 0.01; RR 10.02 [1.23-81.16]). CONCLUSION: Our study is the first to compare the MGS with the BMS in STEMI in the real world, and it also appears to confirm that although the MGS is a safe device in STEMI that is not associated with increased mortality, it is associated with a higher long-term TLR rate.
No takes yet. Share an insight, caveat, or question.
Fernández‐Cisnal et al. (2014) conducted a cohort in ST-elevation myocardial infarction (STEMI) (n=262). MGuard Stent vs. Bare metal stent was evaluated on Target lesion revascularization (TLR) (RR 10.02, 95% CI 1.23-81.16, p=< 0.01). In real-world STEMI patients, the MGuard Stent was associated with a significantly higher rate of target lesion revascularization compared to bare metal stents (11.4% vs 1.3%; RR 10.02; P<0.01).
Synapse has enriched 10 related papers. Consider them for comparative context: