This is the authors' abstract. We don't add key points for this paper.
An Auditory Steady-State Response (ASSR) is a valuable tool for determining auditory thresholds in individuals who are either unable or unwilling to cooperate with conventional behavioral testing methods. This study proposes a sequential test technique for automatic detection of ASSRs, incorporating a non-detection stopping criterion. The electrophysiological thresholds of a normal hearing volunteer were established using data collected from multichannel EEG signals. The detection probabilities and critical values were obtained via Monte Carlo simulations. Remarkably, application of the non-detection stopping criterion resulted in a 60% reduction in exam time in the absence of a response. These findings clearly demonstrate the significant potential of the sequential test in enhancing the performance of automatic audiometry.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Vaz et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: