The estimation of Pioglitazone was done by RP-HPLC.The assay of Pioglitazone was performed with tablets and the % assay was found to be 100.10 which show that the method is useful for routine analysis.Pioglitazone is a medication belonging to the thiazolidinedione class of drugs that are used as adjuncts to diet, exercise, and other diabetes medications to manage type 2 diabetes mellitus.Pioglitazone acts as a selective agonist at Peroxisome Proliferator Activated Receptor Gamma (PPARγ) in target tissues for insulin action such as adipose tissue, skeletal muscle, and liver.UV spectrum of 10 µg/ml Pioglitazone in diluents (mobile phase composition) was recorded by scanning in the range of 200nm to 400nm. the mobile phase was optimized to Acetonitrile: Phosphate buffer (pH 4.5) in proportion 50: 50 v/v respectively.Plot a graph of peak area versus concentration (on Xaxis concentration and on Y-axis Peak area) and calculate the correlation coefficient.Acceptance Criteria: Correlation coefficient should be not less than 0.99.The robustness limit for mobile phase variation and flow rate variation are well within the limit, which shows that the method is having good system suitability and precision under given set of conditions.