Key result
Modified driveline dressing protocol linked to ~91% fewer driveline infections vs previous strategy.
Why the study?
The impact of a standardized driveline care strategy including subfascial-tunneling and dressing protocol on driveline infection incidence in HeartMate II and HeartMate 3 patients was evaluated.
Does a standardized driveline care strategy reduce driveline infection in patients with HeartMate II and HeartMate 3 left ventricular assist devices?
Cohort (n=80)
Does a standardized driveline care strategy reduce driveline infection in patients with HeartMate II and HeartMate 3 left ventricular assist devices?
Absolute Event Rate: 3% vs 35%
p-value: p=0.0009
A standardized triple driveline tunneling strategy and waterproof dressing protocol significantly reduces the incidence of driveline infections in patients with left ventricular assist devices.
May support modified driveline protocols in LVAD patients; hypothesis-generating and requires randomized confirmation.
BACKGROUND: We evaluated the impact of a standardized driveline care strategy, including a subfascial-tunneling method and dressing protocol, on the incidence of driveline infection (DLI). METHODS: DLI data from all HeartMate II (HMII) and HeartMate 3 (HM3) patients (including exchange devices) were retrospectively collected between 2013 and 2021. The driveline subfascial-tunneling method was altered in three steps (A: right direct; B: left triple, C: right triple), and the shower protocol was changed in two steps (A: with/without cover, B: with cover). Disinfection was individually tailored after changing the shower protocol. Complications associated with morbidity and mortality were evaluated for each modification. RESULTS: During the study period, 80 devices were implanted (HMII, n = 54; HM3, n = 26). The 8-year incidence of DLI was 15% (n = 8) in HMII patients and 0% in HM3 patients (p = 0.039). DLI was not associated with hospital mortality. The modified dressing protocol and tunneling method was associated with a significantly better DLI incidence rate in comparison to the previous one: Protocol-A (n = 17), Protocol-B (n = 63), 35% vs 3% (p = 0.0009), Method-A (n = 13), Method-B (n = 42), Method-C (n = 25), 46% vs 5% vs 0% (p = 0.0001). The rete of freedom form DLI at 1, 2, and 3 years had also significant difference between groups: Protocol-A and Protocol-B, 80%, 54%, 54% vs 96%, 96%, 96%, respectively (p < 0.0001), Method-A, Method-B and Method-C, 76%, 44%, 44%, vs 94%, 94%, 94% vs 100%, 100%, respectively (p < 0.0001). CONCLUSIONS: A standardized triple driveline tunneling strategy and waterproof dressing protocol reduced driveline infection in HM3 patients to 0%.
No takes yet. Share an insight, caveat, or question.
Mutsuga et al. (2022) conducted a cohort in Left ventricular assist device (LVAD) implantation (n=80). Modified driveline care strategy (subfascial-tunneling method and dressing protocol) vs. Previous driveline care strategy was evaluated on Incidence of driveline infection (DLI) (p=0.0009). A modified driveline dressing protocol significantly reduced the incidence of driveline infection compared to the previous strategy (3% vs 35%, p=0.0009).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: