Answers are grounded in this paper and cite their sources.
Wastewater-based epidemiology (WBE) was a near-real-time monitoring strategy for illegal drugs. However, solid-phase extraction (SPE) widely used in WBE was time-consuming and labor-intensive to extract ultra-trace target compounds from wastewater. In this study, a convenient magnetic solid-phase extraction (MSPE) approach based on newly designed and synthesized polydopamine functionalized core-shell magnetic mesoporous silica (Fe3O4@nSiO2@mSiO2@PDA) nanocomposite was synthesized and firstly utilized for simultaneously extracting five amphetamine-type stimulants (ATSs) from wastewater samples. Subsequently, ultrahigh performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method coupled with optimal MSPE was developed for determination of ultra-trace ATSs in wastewater. The validation results indicated a favorable linearity ranging from 1 to 200 ng L−1, low limit of detection (0.5–2.5 ng L−1), and qualified recovery (95.1–106.6%) and repeatability (0.6–6.2%). In addition, the Fe3O4@nSiO2@mSiO2@PDA nanoparticles could be reused for at least ten times without significant loss of the adsorption efficiencies of ATSs. Finally, the MSPE-UPLC-MS/MS method was successfully applied to real wastewater samples with the results that the preparation procedure was shrunk from 2 h to 30 min without obvious decline of extraction efficiency compared with the SPE. Hence, based on merits of the novel Fe3O4@nSiO2@mSiO2@PDA nanocomposite, the proposed method is convenient and reliable for determination of ATSs in wastewater.
No takes yet. Share an insight, caveat, or question.
Qi et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: