Key result
Withings BPM Connect meets AAMI/ESH/ISO validation criteria with ~0.6 mmHg systolic difference versus mercury reference.
Why the study?
Electronic devices for blood pressure measurements must undergo independent clinical validations for accuracy assessment as recommended by scientific societies.
Does the Withings BPM Connect device accurately measure blood pressure compared to a mercury sphygmomanometer in the general population?
Observational (n=85)
Does the Withings BPM Connect device accurately measure blood pressure compared to a mercury sphygmomanometer in the general population?
The Withings BPM Connect device meets the AAMI/ESH/ISO Universal Standard accuracy requirements and is validated for home blood pressure measurement in the general population.
Supports Withings BPM Connect for general-population home BP monitoring; leaves open validation in broader or high-risk cohorts.
Background: As recommended by various authorities and scientific societies, electronic devices for blood pressure (BP) measurements must undergo independent clinical validations for accuracy assessment. Objective: To assess the accuracy of the Withings BPM Connect device in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization "AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)". Methods: The Withings BPM Connect device measures BP at the brachial level using the oscillometric method. According to this protocol using the same-arm sequential BP measurement method, subjects (n ≥ 85) fulfilling the age, gender, BP and cuff distribution criteria must be included. Two criteria are used for the analysis: Criterion 1: differences between observers' mercury sphygmomanometer reference measurements and test device BP values (test versus reference) and their standard deviation (SD) must be ≤5 ± 8 mmHg for both systolic (SBP) and diastolic BP (DBP). Criterion 2: the SD of the mean BP differences between the test device and reference BP per subject must be ≤6.91/6.62 mmHg for SBP/DBP. Results: Ninety-two subjects were selected, 85 of whom were included. For validation criterion 1, the mean difference ± SD between the reference and device BP values was 0.6 ± 5.3 mmHg for SBP and 2.1 ± 4.3 mmHg for DBP. For criterion 2, the SD of the mean BP differences between the test device and reference BP per subject was 4.2/3.6 mmHg (SBP/DBP). These results fulfilled the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) protocol requirements. Conclusion: The Withings BPM Connect oscillometric device for home BP measurement fulfilled all of the accuracy requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and consequently can be recommended for home BP measurements.
No takes yet. Share an insight, caveat, or question.
A 2022 study conducted an observational in General population (blood pressure measurement) (n=85). Withings BPM Connect device vs. Mercury sphygmomanometer reference measurements was evaluated on Differences between observers' mercury sphygmomanometer reference measurements and test device BP values (MD 0.6 mmHg (SBP) and 2.1 mmHg (DBP)). The Withings BPM Connect device met AAMI/ESH/ISO validation criteria with a mean difference of 0.6 ± 5.3 mmHg for systolic and 2.1 ± 4.3 mmHg for diastolic blood pressure compared to reference.