Using a large-angle refractometer, the refractive indices of water, methanol, acetonitrile, acetone, ethanol, formaldehyde solution, 2-propanol, hexanes, 1-propanol, 2-butanol, tetrahydrofuran, dichloromethane, 1,4-dioxane, cyclohexane, N,N-dimethylformamide, ethylene glycol, N,N-dimethylacetamide, chloroform, glycerol, trichloroethylene, dimethylsulfoxide, p-xylene, ethylbenzene, toluene, m-xylene, benzene, and o-xylene were measured at 1550 nm and compared to literature values. In addition, the refractive indices of 48 aqueous sucrose solutions, 34 aqueous sodium chloride solutions, 40 DMSO-water, 40 ethylene glycol-water, and 40 glycerol-water mixtures were also measured at different concentrations and fit with third-order polynomial expressions. This study provides near-infrared refractive indices that were not previously available and may help with the calibration of refractive index sensors that operate at or near 1550 nm.
No takes yet. Share an insight, caveat, or question.
Saunders et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: