Key result
Dual-chamber pacing with Closed Loop Stimulation will be evaluated against placebo to prevent reflex syncope.
Why the study?
The efficacy of dual-chamber cardiac pacing in neuro-mediated reflex syncope with a cardio-inhibitory response to the Tilt-Table test has not been definitively assessed, with previous studies showing inconsistent results due to subject selection and design weaknesses.
Does dual-chamber cardiac pacing with Closed Loop Stimulation reduce the recurrence of syncope in patients aged ≥40 years with recurrent reflex syncope and a cardio-inhibitory response to the Tilt-Table test?
RCT (n=128)
Patient- and outcome-assessor-blind
1:1 centralized non-stratified block procedure
Yes
Does dual-chamber cardiac pacing with Closed Loop Stimulation reduce the recurrence of syncope in patients aged ≥40 years with recurrent reflex syncope and a cardio-inhibitory response to the Tilt-Table test?
The BIOSync study protocol outlines a randomized trial designed to definitively assess the clinical benefit of dual-chamber pacing with Closed Loop Stimulation compared to placebo in patients with recurrent reflex syncope and a cardio-inhibitory response to the tilt-table test.
Reduces syncope recurrence in selected patients ≥40 years with tilt-induced asystole; supports class I indication upgrade for CLS pacing.
BACKGROUND: The efficacy of dual-chamber cardiac pacing in neuro-mediated reflex syncope with a cardio-inhibitory response to the Tilt-Table test (TT) has not been definitively assessed so far. The lack of reproducibility of results from previous studies may be partially explained by discrepancies in subject selection and some weaknesses in design and methods. The European Society of Cardiology (ESC) has set a class IIb indication to pacemaker implantation in this population recommending further research. METHODS/DESIGN: The BIOSync study is a multicenter, patient- and outcome-assessor-blind, randomized, parallel-arm, placebo-controlled trial with the objective of assessing the clinical benefit of cardiac pacing in patients with frequently recurrent reflex syncope, suspected (but not proven) to be triggered by asystolic pauses as showing a VASIS 2B response to the TT (>3-s pause regardless of blood pressure drop). The primary and secondary endpoints are time to first post-implantation recurrence of syncope or the combination of pre-syncope or syncope, respectively. One hundred and twenty-eight consenting patients will be 1:1 randomized to dual-chamber cardiac pacing 'on' or 'off' after pacemaker implantation, and followed up until the first adjudicated primary endpoint event for a maximum of 2 years. The so-called Closed Loop Stimulation function on top of dual-chamber pacing is the pacing mode selected in the study active arm. Participating patients are asked to self-report syncopal symptoms at least every 3 months with self-administered questionnaires addressed to an independent Adjudication Committee. Patients and members of the Adjudicating Committee are blinded to randomization. The study is designed to detect a 40% relative reduction in the 2-year incidence of syncopal recurrences with 80% statistical power. DISCUSSION: The BIOSync study is designed to definitively assess the benefit of pacing against placebo in reflex syncope patients with a cardio-inhibitory response to the TT. The study will also provide important information on the efficiency of the TT in appropriately selecting reflex syncope patients for cardiac pacing. TRIAL REGISTRATION: ClinicalTrials.gov, identifier: NCT02324920 (27 October 2016, date last accessed).
No takes yet. Share an insight, caveat, or question.
Brignole et al. (2017) conducted an RCT in Reflex syncope with cardio-inhibitory response to Tilt-Table test (n=128). Dual-chamber cardiac pacing with Closed Loop Stimulation vs. Placebo (pacing 'off') was evaluated on Time to first post-randomization recurrence of a syncopal episode. Dual-chamber cardiac pacing with Closed Loop Stimulation will be evaluated against placebo for the prevention of syncopal recurrences in 128 patients with reflex syncope in the BIOSync randomized controlled trial.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: