Key result
Temporary epicardial pacing linked to two VF episodes due to R-on-T from PVC undersensing.
Why the study?
Undersensing of epicardial pacing wires may cause the R-on-T phenomenon leading to ventricular fibrillation, but the mechanisms and prevention are not well understood.
Case Report (n=1)
No
Epicardial pacemakers may undersense wide QRS complexes like PVCs despite low sensing thresholds, potentially leading to dangerous R-on-T phenomenon and ventricular fibrillation.
Epicardial pacing may undersense wide QRS leading to R-on-T VF; this case leaves open optimal sensing thresholds.
Epicardial pacemaker wire insertion is standard following cardiothoracic surgery. However, undersensing of pacing wires may cause the R-on-T phenomenon, which induces ventricular fibrillation. We report a case of a male patient with severe mitral regurgitation scheduled for mitral valve replacement who experienced two ventricular fibrillation episodes related to the R-on-T phenomenon caused by undersensing of the epicardial pacing wire. Both undersensing events happened despite an appropriately low sensing threshold. Notably, the stimulated T wave followed the QRS of the premature ventricular contraction (PVC). This case suggests that a PVC's R wave may be undersensed despite a low sensing threshold. This critical complication may have occurred because pacemakers sense R waves using a slew rate, which is the quotient of voltage over time. As a result, pacemakers may undersense wide QRS waves such as PVCs. Avoiding this dangerous phenomenon completely is not possible using epicardial pacemakers; therefore we recommend carefully adapting epicardial pacing especially when PVC waves occur frequently.
No takes yet. Share an insight, caveat, or question.
Nakamori et al. (2016) conducted a case report in Ventricular fibrillation (n=1). Temporary epicardial pacing was evaluated on Ventricular fibrillation episodes. Temporary epicardial pacing caused two episodes of ventricular fibrillation due to the R-on-T phenomenon from undersensing of premature ventricular contractions despite a low sensing threshold.