We report the case of a 48-year-old male who was found to have an atrial septal defect (ASD) of the secundum type with evidence of moderate right heart load requiring elective device closure. Intraprocedural balloon-sizing showed an average stretched diameter of 16 mm. The decision was made to implant a 17 mm AMPLATZER® Septal Occluder (St Jude Medical, St Paul, MN, USA). The left-sided countercluder has a diameter of 31 mm, making sizing potentially critical as the patient has a normal left atrial (LA) size. The device implantation was performed uneventfully using intracardiac echocardiographic (ICE) guidance. ICE confirmed adequate deployment but showed the left-sided countercluder to be in contact with the aortic wall and the LA roof. Contrast cardiac computed tomography was carried out the day after the procedure and provided data for subsequent three-dimensional (3D) printing in end-diastole (Materialise Inc., Leuven, Belgium). The virtual and tangible 3D printouts demonstrated the device to adequately fill the ASD and both countercluders to not interfere with venous inlets and to be perfectly aligned with the interatrial septum (Panels A and B show virtual printouts and Panels C and D real printouts). The left-sided countercluder was shown to contact but not to significantly indent the aortic wall or the LA roof (Panels E and F show real printouts).