Answers are grounded in this paper and cite their sources.
Objective: The study presents the validation protocol for the determination of total phosphorus content at parts per million (ppm) (μg/L) levels in animal feed by a UV-Vis spectrophotometer. Methods: The measured absorbance of solutions against the blank solution was at 400 nm with the spectrophotometer. A combined ammonium heptamolybdate tetrahydrate and ammonium monovanadate blue solution was used as a coloring reagent for detection. This method was validated by evaluation of statistical parameters such as linearity, sensitivity, limits of detection (LOD) and quantification (LOQ), precision, accuracy, and measurement uncertainty using a matrix blank (MB) against the phosphorus standard. Results: The Instrumental Detection Limit was 0.066 ppm and the Instrumental Quantification Limit was 0.22 ppm, respectively, while the phosphorus recovery and repeatability percent were 101.15% and 0.11%, respectively. However, the linearity of this method was 0.1 to 30 ppm. The measurement uncertainty of this method was 2.82%, following Commission Regulation (EC). Conclusion: The estimated parameters in the validation protocol, were found to meet the imposed performance criteria, and the procedure was validated for the intended use.
No takes yet. Share an insight, caveat, or question.
Debnath et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: