This is the authors' abstract. We don't add key points for this paper.
An ultrasensitive electrochemical immmunosensor was demonstrated to be capable of detecting the hormone cortisol down to concentrations as low as 16 pg mL(-1). In addition, the immunosensor displayed a sensitivity of 1.6 μA pg(-1) mL(-1) and a linear range up to ∼2500 pg mL(-1) of cortisol. This immunosensor was constructed based on a Au nanoparticle|dimethyl 3,3'-dithiobispropionimidate·2HCl (DTBP)-Protein G scaffold-modified Au electrode. In this work, the Au nanoparticles were used to increase the electrochemically active surface area by 28% (with a standard deviation of 3%) to enhance the quantity of the Protein G scaffold on the electrode. Thiolation of Protein G by DTBP aided in avoiding the confirmation change of Protein G, while this Protein G-DTBP component offered an orientation-controlled immobilisation of the capture antibody on the Au electrode. In this immunosensor, a monoclonal anti-cortisol capture antibody was optimally aligned by the scaffold before a competitive immunoassay between sample cortisol and a horseradish peroxidase-labelled cortisol conjugate was conducted. For quantitative analysis, square wave voltammetry was used to monitor the reduction current of benzoquinone produced from a horseradish peroxidase catalysed reaction. The improved analytical performance of our immunosensor was attributed to the synergetic effect of Au nanoparticles and the Protein G-DTBP scaffold.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: