Key result
Rapid calf-IPC produces the greatest hemodynamic effects versus baseline, while calf NMES shows unacceptable tolerability.
Why the study?
An optimized treatment modality for stimulation of venous return to prevent venous thromboembolism has yet to be defined.
Do lower-limb intermittent pneumatic compression devices and neuromuscular stimulation improve venous hemodynamics compared to baseline in healthy individuals?
Do lower-limb intermittent pneumatic compression devices and neuromuscular stimulation improve venous hemodynamics compared to baseline in healthy individuals?
Effect estimate: mean effect size 2.06
Calf-compressing devices provide superior venous hemodynamic augmentation compared to foot devices, but neuromuscular electrical stimulation is limited by poor tolerability.
Stimulation of venous return is critical for prevention of venous thromboembolism; however, an optimized treatment modality has yet to be defined. Therefore, the main objective of this study was to investigate and compare the hemodynamic effects and the second aim was to assess tolerability of three different intermittent pneumatic compression (IPC) devices, foot-IPC, rapid, and slow calf-IPC, and the recently developed technology of calf neuromuscular electric stimulation (NMES). The blood flow in the right popliteal vein was assessed with Doppler ultrasound in 10 healthy individuals in semirecumbent position without (baseline) and with each of the IPC- and NMES treatment devices. Comfort assessment was completed using visual analogue scale for each treatment modality. All treatments significantly increased peak velocity (PV), mean velocity (MV) volume flow (VF) and ejected volume per individual stimulus (EV) as compared to baseline. Rapid calf-IPC caused the most profound hemodynamic effects with a mean effect size 2.06, taking into account all the above hemodynamic parameters compared to baseline. The weakest response was observed with foot-IPC with a mean effect size 1.50 compared to baseline. All devices compressing the calf produced higher hemodynamic values (mean effect size: 1.89 slow calf-IPC and 1.58 NMES) compared to foot-IPC. Comfort assessment demonstrated unacceptable tolerability for NMES, while the other devices showed an acceptable tolerability. Augmentation of venous return using calf- compared to foot devices is hemodynamically superior. Varying hemodynamic responses are observed between calf blood flow stimulating devices. The low tolerability of NMES, however, may limit its clinical applicability.
No takes yet. Share an insight, caveat, or question.
Praxitelous et al. (2018) studied Healthy individuals (n=10). Intermittent pneumatic compression (IPC) devices and calf neuromuscular electric stimulation (NMES) vs. Baseline (no device) was evaluated on Hemodynamic effects (peak velocity, mean velocity, volume flow, and ejected volume) (mean effect size 2.06). Rapid calf-IPC produced the most profound hemodynamic effects (mean effect size 2.06 vs baseline), while calf neuromuscular electric stimulation demonstrated unacceptable tolerability.