Key result
BD Whitacre 27G needles retain less residual volume after spinal anesthesia versus Quincke and disposable needles.
Why the study?
The volume of residual fluid remaining inside spinal needles after spinal anesthesia administration varies by needle design and has not been quantitatively compared across different 27 gauge needles.
Does the BD Whitacre pencil point 27G needle reduce residual volume compared to Spinocan Quincke and Disposable spinal needles after spinal anesthesia?
Cross-Sectional
Does the BD Whitacre pencil point 27G needle reduce residual volume compared to Spinocan Quincke and Disposable spinal needles after spinal anesthesia?
p-value: p=<0.01
BD Whitacre pencil point 27G needles lodge less residual volume following spinal anesthesia compared to other brands, though clinical implications remain unclear.
Whitacre 27G needles retain less residual volume; leaves open any dosing or clinical impact in spinal anesthesia.
Introduction Spinal anesthesia involves the administration of a dose of local anesthesia into the intrathecal space of the lumbar spine, using spinal needles with varying characteristics and design. When completing the administration of spinal anesthesia, you may check the volume of fluid remaining inside the spinal needle. Objective To measure and compare the volume of residual fluid following the administration of spinal anesthesia using three different brands of 27 gauge spinal needles. Methods This is an observational analytical cross-section trial for which tree types of 27 gauge commercial needles were selected: BD Whitacre pencil point, Spinocan Quincke sharp point, and Disposable spinal needle. Following the administration of spinal anesthesia, the same operator, without detaching the syringe from the needle, deposited the residual volume into a collection tube so that a second operator quantifies the volume using a precision micropipette. Results The residual volume obtained from the 27G spinal needles showed a statistically significant difference in favor of the BD Whitacre needles ( P < 0.01), compared against Spinocan Quincke and Disposable spinal needle. Conclusions The BD Whitacre pencil point proved to lodge less residual volume following spinal anesthesia. No clinical implications can be made from this finding and new studies are required to ratify these results.
No takes yet. Share an insight, caveat, or question.
Luna et al. (2017) conducted a cross-sectional in Spinal anesthesia. BD Whitacre pencil point 27G spinal needle vs. Spinocan Quincke sharp point and Disposable spinal needle was evaluated on Residual volume following the administration of spinal anesthesia (p=<0.01). BD Whitacre pencil point 27G spinal needles lodged significantly less residual volume following spinal anesthesia compared to Spinocan Quincke and Disposable spinal needles (P<0.01).