Key result
Pencil-point spinal needles exhibited the least deviation (0.60-1.00 mm/50 mm tissue) compared to epidural and beveled spinal needles when passed through porcine tissue (P<0.001).
Why the study?
Does needle point design and gauge influence the amount of needle deviation in a porcine tissue model?
Does needle point design and gauge influence the amount of needle deviation in a porcine tissue model?
p-value: p=<0.001
Needle point design and gauge significantly influence needle deviation in tissue, with pencil-point spinal needles exhibiting the least deviation.
May favor pencil-point needles to limit deviation in neuraxial procedures; leaves open human validation and outcome trials.
Although many characteristics of needles used for spinal and epidural anesthesia have been studied extensively, the amount that a needle deviates from a straight path while passing through tissue has been relatively ignored. A laboratory model was used to determine the amount of needle deviation produced when different types of new spinal needles and epidural needles were passed through porcine tissue. Both needle point design and, to a lesser extent, needle gauge were found to influence the amount and direction of needle deviation. Deviation was found to be the least for the pencil-point spinal needles (Whitacre, Sprotte, Safetap, range 0.60-1.00 mm/50 mm tissue, analysis of variance (ANOVA) P < 0.001). Needles commonly used for epidural anesthesia (Tuohy, Hustead, Crawford) exhibited increased deviation (range 1.73-3.54 mm/50 mm tissue), although the largest amount was seen with beveled spinal needles (Quincke, Atraucan) (range 4.42-5.90 mm/50 mm tissue). The possible clinical significance of needle deviation during the performance of a regional anesthetic is discussed.
No takes yet. Share an insight, caveat, or question.
Kopacz et al. (1995) studied Regional anesthesia. Pencil-point spinal needles vs. Epidural and beveled spinal needles was evaluated on Amount of needle deviation per 50 mm tissue (p=<0.001). Pencil-point spinal needles exhibited the least deviation (0.60-1.00 mm/50 mm tissue) compared to epidural and beveled spinal needles when passed through porcine tissue (P<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: