Key result
Fully automated closed-loop IABP achieves beneficial diastolic augmentation without manual intervention in ~80% of patients.
Why the study?
A new generation closed loop, fully automatic intraaortic balloon pump system was developed to optimize diastolic augmentation without operator intervention and its clinical suitability needed investigation.
Does a closed loop, fully automatic intra-aortic balloon pump provide safe and effective diastolic augmentation without operator intervention in patients requiring IABP?
Does a closed loop, fully automatic intra-aortic balloon pump provide safe and effective diastolic augmentation without operator intervention in patients requiring IABP?
A fully automated, closed-loop IABP system is clinically feasible and provides consistent, safe diastolic augmentation without requiring continuous operator intervention.
Should not change practice after 10 cases; leaves open need for randomized trials of automated IABP.
A new generation, closed loop, fully automatic intraaortic balloon pump (CL-IABP) system continuously optimizes diastolic augmentation by adjusting balloon pump parameters beat by beat without operator intervention. In dogs in sinus rhythm and with experimentally induced arrhythmias, the new CL-IABP system provided safe, effective augmentation. To investigate the system's suitability for clinical use, 10 patients meeting standard indications for IABP were studied. The patients were pumped by the fully automatic IABP system for an average of 20 hr (range, 1-48 hr). At start-up, the system optimized pumping parameters within 7-20 sec. Evaluation of 186 recordings made at hourly intervals showed that inflation began within 20 msec of the dicrotic notch 99% of the time. In 100% of the recordings, deflation straddled the first half of ventricular ejection. Peak pressure across the balloon membrane averaged 55 mmHg and, in no case, exceeded 100 mmHg. Examination of the data showed that as soon as the system was actuated it provided consistently beneficial diastolic augmentation without any further operator intervention. Eight patients improved and two died (one of irreversible cardiogenic shock and one of ischemic cardiomyopathy). No complications were attributable to the investigational aspects of the system. A fully automated IABP is feasible in the clinical setting, and it may have advantages relative to current generation IABP systems.
No takes yet. Share an insight, caveat, or question.
Kantrowitz et al. (1992) studied Standard indications for intra-aortic balloon pump (IABP) (n=10). Closed loop, fully automatic intraaortic balloon pump (CL-IABP) system was evaluated on Inflation began within 20 msec of the dicrotic notch. A fully automated closed-loop intra-aortic balloon pump provided beneficial diastolic augmentation without operator intervention in 10 patients, with 8 improving and 2 dying.