Key result
Transtracheal Doppler cardiac output correlates closely with thermodilution with an R² of ~0.84.
Why the study?
Transtracheal Doppler was developed to provide an estimate of cardiac output in endotracheally intubated humans, requiring validation against thermodilution.
Does transtracheal Doppler provide accurate continuous cardiac output measurements compared to thermodilution in intubated patients?
Observational
Does transtracheal Doppler provide accurate continuous cardiac output measurements compared to thermodilution in intubated patients?
Transtracheal Doppler incorporated into an endotracheal tube provides cardiac output estimates that correlate well with thermodilution in intubated patients.
Answers are grounded in this paper and cite their sources.
May enable cardiac output estimation in intubated patients; leaves open prospective validation before clinical adoption.
Transtracheal Doppler, a procedure developed in the authors' laboratory, provides an estimate of cardiac output in endotracheally intubated dogs. The present study reports initial results in humans with a Doppler probe incorporated into an endotracheal tube. The system was first calibrated by empirically determining phi, the angle of the ultrasound beam with respect to the blood velocity vector. For a best least squares fit for phi, aortic diameters can be calibrated over a range of 22-37 mm. The calibration was then tested in a separate group of patients using the same empirically derived angle to calculate cardiac outputs. The transtracheal Doppler cardiac outputs show good correlation with those obtained by thermodilution over a range of cardiac outputs, 2.69-8.62 1/min, R2 = 0.835.
No takes yet. Share an insight, caveat, or question.
Abrams et al. (1989) conducted an observational in Endotracheally intubated patients. Transtracheal Doppler vs. Thermodilution was evaluated on Cardiac output (R2 = 0.835). Transtracheal Doppler cardiac outputs showed good correlation with those obtained by thermodilution over a range of 2.69-8.62 L/min.