Key result
ECG-guided LV lead placement linked to ~70% higher CRT echocardiographic response rates vs unguided implantation.
Why the study?
Failure to select the optimal left ventricular segment for lead implantation is a major cause of unresponsiveness to cardiac resynchronization therapy.
Does intraoperative surface ECG-guided left ventricular lead placement improve echocardiographic and clinical response in heart failure patients undergoing CRT-D implantation?
Cohort (n=80)
Does intraoperative surface ECG-guided left ventricular lead placement improve echocardiographic and clinical response in heart failure patients undergoing CRT-D implantation?
Absolute Event Rate: 85% vs 50%
p-value: p=0.02
Intraoperative surface ECG-guided LV lead placement targeting the shortest biventricular-paced QRS duration significantly improves echocardiographic response to CRT compared to standard unguided placement.
Should not yet change CRT lead placement practice; leaves open need for randomized confirmation of response benefit.
OBJECTIVE: Failure to select the optimal left ventricular (LV) segment for lead implantation is one of the most important causes of unresponsiveness to the cardiac resynchronization therapy (CRT). In our study, we aimed to investigate the echocardiographic and clinical benefits of LV lead implantation guided by an intraoperative 12-lead surface electrocardiogram (ECG) in patients with multiple target veins. METHODS: We included 80 [42 (62.5%) male] heart failure patients who successfully underwent CRT defibrillator (CRT-D) implantation. Patients were divided into two groups. In group 1, LV lead was positioned at the site with the shortest biventricular-paced (BiV-paced) QRS duration (QRSd), as intraprocedurally measured using surface ECG. In group 2 (control), we included patients who underwent the standard unguided CRT. ECG, echocardiogram, and functional status were evaluated before and 6 months after CRT implantation in all patients. RESULTS: In group 1, BiV-paced QRSd measurements were successfully performed in 112 of 120 coronary sinus branches during CRT and an LV lead was successfully placed at the optimal site in all patients. Compared with group 2, group 1 had a significantly higher rate (85% vs. 50%, p=0.02) of response (>15% reduction in LV end-systolic volume) to CRT as well as a shorter QRSd (p<0.001) and a greater QRS shortening (ΔQRS) associated with CRT compared with baseline (p<0.001). The mean New York Heart Association functional class was significantly improved in both groups, and no significant differences were found in clinical response to CRT (85% vs. 70%, p=0.181). CONCLUSION: Surface ECG can be used to guide LV lead placement in patients with multiple target veins for improving response to CRT. Thus, it is a safe, feasible, and economic approach for CRT-D implantation.
No takes yet. Share an insight, caveat, or question.
Abdülcebbar Şipal (2018) conducted a cohort in Heart failure (n=80). Intraoperative 12-lead surface ECG-guided left ventricular lead placement vs. Standard unguided CRT was evaluated on Response to CRT (>15% reduction in LV end-systolic volume) (p=0.02). Intraoperative surface ECG-guided left ventricular lead placement significantly increased the rate of echocardiographic response to CRT compared to standard unguided placement (85% vs. 50%, p=0.02).