Key result
Suturing reduces bovine pericardial tissue strength by up to ~59%, driven by larger needle sizes.
Why the study?
The mechanical impact of suturing on bovine pericardial tissue used in bioprosthetic heart valve fabrication was not well characterized.
Does suturing weaken the mechanical properties of bovine pericardial tissue used for bioprosthetic heart valves?
Does suturing weaken the mechanical properties of bovine pericardial tissue used for bioprosthetic heart valves?
Suturing significantly weakens bovine pericardial tissue, highlighting a critical mechanical consideration in the fabrication of bioprosthetic heart valves.
Suturing weakens pericardial tissue; leaves open optimal techniques to preserve durability in bioprosthetic valves.
In the fabrication of a bioprosthetic heart valve from bovine pericardial tissues, the tissues are subjected to suturing. The stress-strain response of sutured as well as unsutured strips of this tissue were examined. The stress-strain response was determined using a tensile-testing machine. It was found that suturing weakens the tissue in that sutured strips are more extensible, exhibit a lower stress at rupture and a lower final elastic modulus. In addition, it was also found that the bigger the suture/needle size used the greater the decrease in tissue strength. In all, tissue strength was observed to decrease by 22 to 59% in this study. The weakening of the tissue is attributed to the puncture holes created by the surgeon's needle which create regions of weakness. This response of bovine pericardial tissue to suturing should be given due consideration in the fabrication of a bioprosthetic heart valve using this tissue.
No takes yet. Share an insight, caveat, or question.
Lim et al. (1994) studied this question. Suturing vs. Unsutured strips was evaluated on Tissue strength (stress-strain response) (22 to 59% decrease). Suturing bovine pericardial tissue decreased tissue strength by 22 to 59% compared to unsutured strips, with larger suture/needle sizes causing greater decreases in strength.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: