Key result
Automated CytoCamTools shows acceptable mean bias of ~2.2 mm/mm² for vessel density but wide agreement limits.
Why the study?
Analysis of microvascular density parameters is time consuming and operator-dependent, limiting its use as a point-of-care tool in clinical practice.
Cross-Sectional (n=22)
An automatic analysis software has been developed to quickly detect microvascular density parameters, potentially enabling point-of-care microvascular monitoring.
Answers are grounded in this paper and cite their sources.
Automatic software may enable point-of-care microvascular monitoring in the ICU; leaves open prospective outcome validation.
Analysis of microvascular density parameters is time consuming and operator-dependent. 1 This is the main limitation to use microvascular monitoring in clinical practice as a “point-of-care” tool. Recently, an automatic analysis software has been developed and could allow us to obtain results quickly.
No takes yet. Share an insight, caveat, or question.
Carsetti et al. (2015) conducted a cross-sectional in Post-cardiac surgery (n=22). CytoCamTools 1.7.12 automatic analysis software vs. Automated Vascular Analysis (AVA) software 3.2 by skilled operators was evaluated on Mean bias in total vessel density (TVD) between CC and AVA (Mean bias 2.17 mm/mm^2, 95% CI 1.44 to 2.90, p=0.0001). The automated CytoCamTools software showed an acceptable mean bias of 2.17 mm/mm^2 for total vessel density compared to skilled operators using AVA, but with wide limits of agreement.