Key result
SavvyWire delivers effective rapid pacing with hemodynamic drop in ~98% of TAVR patients.
Why the study?
The efficacy and safety of the SavvyWire guidewire during TAVR procedures had not been established.
Does a guidewire with a pressure sensor and pacing properties provide effective rapid pacing and safety in patients with severe aortic stenosis undergoing TAVR?
Observational (n=121)
Yes
Does a guidewire with a pressure sensor and pacing properties provide effective rapid pacing and safety in patients with severe aortic stenosis undergoing TAVR?
Answers are grounded in this paper and cite their sources.
The use of a dedicated pressure sensor and pacing guidewire during TAVR is safe and highly effective for achieving rapid pacing and continuous hemodynamic monitoring without additional catheter exchanges.
BACKGROUND: The SavvyWire (OpSens Inc) is a 0.035-inch preshaped guidewire with dedicated pacing properties and a distal pressure sensor allowing for continuous hemodynamic pressure monitoring. OBJECTIVES: This study sought to determine the efficacy and safety of the guidewire during transcatheter aortic valve replacement (TAVR) procedures. METHODS: This prospective, multicenter clinical study included patients with severe aortic stenosis undergoing TAVR in 8 European centers. The primary efficacy endpoint was defined as effective left ventricular rapid pacing runs with the guidewire translating into a significant systemic pressure drop (below 60 mm Hg). The safety outcome included the absence of major procedural complications related to the guidewire. RESULTS: A total of 121 patients (mean age: 82.2 ± 5.9 years, 50% women) were included in the study, and 119 (98.3%) patients were finally treated with the study device. A balloon-expandable valve was implanted in 45 (37.8%) patients. Predilatation and postdilatation were performed in 89 (74.8%) and 14 (11.8%) patients, respectively. The primary efficacy endpoint was achieved in 116 (98.3%) patients, and the mean aortic systolic arterial pressure achieved during rapid pacing was 46.6 ± 11.3 mm Hg. Hemodynamic assessment with the use of the OptoMonitor 3 (OpSens Inc) without additional catheter exchange was achieved in 117 (99.2%) patients. The safety endpoint was achieved in 117 (99.2%) patients. No procedural mortality, stroke, or ventricular perforation was reported. CONCLUSIONS: The use of the guidewire during TAVR procedures appeared to be efficacious and safe. This device could help minimize interventions during the procedure and improve the clinical decision making after transcatheter heart valve deployment. (SavvyWire Efficacy and Safety in Transcatheter Aortic Valve Implantation Procedures [SAFE-TAVI]; NCT05492383).
No takes yet. Share an insight, caveat, or question.
Regueiro et al. (2023) conducted an observational in Severe aortic stenosis (n=121). SavvyWire pressure sensor and pacing guidewire was evaluated on Effective left ventricular rapid pacing runs translating into a significant systemic pressure drop (below 60 mm Hg). The SavvyWire pressure sensor and pacing guidewire achieved effective left ventricular rapid pacing with a significant systemic pressure drop in 98.3% of patients undergoing TAVR.