Key result
EVOQUE transcatheter tricuspid valve replacement in patients with pre-existing transvenous RV leads resulted in procedure-related lead complications in 4 of 52 patients (7.7%).
Why the study?
Outcomes of patients with pre-existing transvenous right ventricular leads undergoing EVOQUE transcatheter tricuspid valve replacement and lead entrapment were not well characterized.
Cohort (n=52)
Yes
EVOQUE TTVR in patients with pre-existing RV leads is associated with a low rate of major lead-related complications, though changes in lead parameters may occur.
Answers are grounded in this paper and cite their sources.
Supports cautious EVOQUE TTVR consideration with pre-existing RV leads; leaves open need for prospective lead outcome data.
BACKGROUND: In patients with a cardiac implantable electronic device (CIED) using a transvenous right ventricular (RV) lead, transcatheter tricuspid valve replacement (TTVR) results in RV lead entrapment. OBJECTIVES: The authors sought to evaluate outcomes of patients with pre-existing transvenous RV leads undergoing EVOQUE TTVR. METHODS: Patients with a CIED referred for TTVR at 3 centers were evaluated preoperatively by the heart team. Lead and clinical assessments were performed at baseline and in follow-up per local clinical practice. The primary outcome was the nature and incidence of major lead-related complications after TTVR. Secondary outcomes included changes in lead parameters after entrapment. RESULTS: Of 146 EVOQUE TTVR, 52 consecutive patients with entrapped RV leads (age 81 [77-85] years; 33 female patients (63.5%), lead dwell time 8.9 [3.3-13.6] years) were evaluated. Paravalvular leak was mild or less in 47 of 48 patients (98%) 30 days post-TTVR. Four patients had procedure-related lead complications: 1 intraprocedural RV lead dislodgement; 2 implantable cardioverter-defibrillator lead fractures 6- and 8-weeks post-TTVR; and 1 methicillin-resistant Staphylococcus aureus endocarditis 1 month post-TTVR. Among 46 complication-free patients with follow-up >30 days (356 [196-965] days), there were no other lead-related adverse events. Of the 33 patients without complications who had ≥1 CIED interrogation >30 days (424 [182-1,006] days) following TTVR, there was a significant decrease in lead impedance and increase in capture threshold on the final lead assessment, but no additional patients required lead revision. CONCLUSIONS: Following TTVR, most with entrapped RV leads remain free of lead-related complications. The need for urgent CIED revision due to RV lead dislodgment or failure is uncommon but may occur.
No takes yet. Share an insight, caveat, or question.
Peigh et al. (2025) conducted a cohort in Pre-existing transvenous right ventricular leads undergoing transcatheter tricuspid valve replacement (n=52). EVOQUE transcatheter tricuspid valve replacement (TTVR) was evaluated on Nature and incidence of major lead-related complications after TTVR. EVOQUE transcatheter tricuspid valve replacement in patients with pre-existing transvenous RV leads resulted in procedure-related lead complications in 4 of 52 patients (7.7%).