This is the authors' abstract. We don't add key points for this paper.
Urea level determination in biological fluids provides nutritional information regarding the body metabolism and/or the renal condition. In order to propose a rapid determination of this biomarker at low levels in biological fluid samples, the present work developed an electrochemical sensor based on a simple arrangement of a nickel-nanoparticle-modified glassy carbon electrode (GCE/NiNPs). Under optimal conditions for selective urea detection, a linear response was obtained in the range of 0.085 to 3.10 mmol L−1 (R = 0.9993), with a limit of detection of 60.0 µmol L−1 and limit of quantification of 198.0 µmol L−1. The GCE/NiNPs sensor showed recovery rates from samples from 105 ± 0.2% (hemodialysis wastewater) to 111 ± 0.3% (dialysate) for urea determination in biological fluid samples. The results obtained showed that the proposed method is relatively easy to operate and has a good degree of reliability and the satisfactory sensitivity required for monitoring urea, as well as exhibiting excellent detection capacity in the presence of possible interferents in both samples.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Oliveira et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: