A sensitive reversible optical chemical sensor for detecting lead ions is described. The sensor is based on the interaction of Pb(II) with diphenylcarbazone in plasticized PVC membrane incorporating a plasticizer (tributylphosphate). The membranes were cast onto glass substrates and used for the determination of lead ions in aqueous solutions by batch and flow-through methods. When the sensing film came in contact with lead(II) ions at pH 6.0, its color changed from pale orange to pink. Parameters affecting the optode response such as pH, type of plasticizer, plasticizer/PVC ratio, and diphenylcarbazone percent were studied. The optode was found to have a long linear dynamic range, 6 9×10−6 to 1 1×10−2 mol L−1 Pb(II), and a short response time of 180 s with a detection limit of 6 5×10−6 mol L−1 Pb(II) by batch spectrophotometric method. In addition, the sensor has been incorporated into a home-made flow-through cell for determining lead ions in flowing streams with improved sensitivity, precision, and detection limit. In the flow system, the linear calibration range was 1 0×10−6 to 5 0×10−3 mol L−1 Pb(II). The optode could be recovered several times after treating with a 0.005 mol L−1 nitric acid solution. Under the experimental conditions, the response of the optode was fully reversible. The effect of the sample matrices on the determination of Pb(II) was studied. The sensor was used for the determination of Pb(II) ions in river water, tap water, and industrial waste samples with satisfactory results.