Answers are grounded in this paper and cite their sources.
Purpose: The aim of this study was to assess the accuracy of ultra-low dose (ULD) CBCT in orthodontic landmark identification compared with standard dose CBCT on reconstructed lateral and posteroanterior views. Materials and Methods: 10 dry human skulls were scanned twice; once with standard CBCT standard radiation dose, and with ULD CBCT protocol operated at 90 kVp, 7.1 mA, 4.5 sec and 400µm voxel size. From each scan, a lateral and a posteroanterior cephalogram were reconstructed, and 33 landmarks were identified on them by 2 experts. All measurements were then repeated after 2 weeks for intra-observer reliability. Results: There was no statistically significant difference in landmarks identification between ULD and standard dose CBCT. The mean difference between both groups ranged between -0.12 to 0.51 in lateral view and -0.67 to 0.46 in posteroanterior view with excellent inter and intra observer agreement with an ICC value >0.7. Conclusion: Ultra-low dose CBCT with reconstructed lateral and posteroanterior views may be recommended for regular orthodontic diagnostic measurements and can replace the standard dose CBCT
No takes yet. Share an insight, caveat, or question.
Abbady et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: