Key result
Novel sensorized heart valve detects altered leaflet motion via a ~29% drop in impedance variation.
Why the study?
Subclinical valve thrombosis in heart valve prostheses can cause reduced leaflet motion detectable only by advanced imaging, motivating development of earlier detection methods.
Does a novel sensorized heart valve prosthesis detect altered leaflet motion in an in vitro circulatory mock loop system?
Does a novel sensorized heart valve prosthesis detect altered leaflet motion in an in vitro circulatory mock loop system?
Absolute Event Rate: 16.7% vs 23.4%
p-value: p=<0.05
A novel sensorized heart valve prosthesis can detect altered leaflet motion in vitro, showing potential for early warning of subclinical valve thrombosis.
Supports in vitro feasibility of sensorized valve monitoring; hypothesis-generating for early thrombosis detection pending clinical validation.
BACKGROUND: Recent studies have shown that subclinical valve thrombosis in heart valve prosthesis (HVP) can be responsible for reduced leaflet motion detectable only by advanced imaging diagnostics. We conceived a novel sensorized HVP able to detect earlier any thrombus formation that may alter the leaflets motion using an electric impedance measurement, IntraValvular Impedance (IVI). METHODS: For IVI measurement, dedicated electrodes are embedded in the structure of the HVP to generate a local electric field that is altered by the moving valve leaflets during their cyclic opening/closing. We present preliminary in vitro results using a first prototype of sensorized mechanical heart valve connected to an external impedance measurement system. The prototype was tested on a circulatory mock loop system and the IVI signals were recorded during both normal dynamics and experimentally induced altered working of the leaflets. RESULTS: Recordings showed a very repetitive and stable IVI signal during the normal cyclic opening/closing of the HVP. The induced alterations in leaflet motion were reflected in the IVI signal. CONCLUSIONS: The novel sensorized HVP has great potential to give early warning of possible subclinical valve thrombosis altering the valve leaflet motion, and to help in tailoring the anticoagulation therapy.
No takes yet. Share an insight, caveat, or question.
Marcelli et al. (2018) studied Heart valve prosthesis thrombosis. Sensorized heart valve prosthesis with IntraValvular Impedance (IVI) measurement vs. Normal leaflet dynamics was evaluated on Maximum percent variation of the impedance module (ΔIVIMAX%) (p=<0.05). The novel sensorized heart valve prosthesis successfully detected experimentally induced alterations in leaflet motion, showing a significant reduction in maximum impedance variation (16.7% vs 23.4%, p<0.05).