Key result
Extravascular ICDs demonstrate ~89% freedom from major system or procedure complications at one year.
Why the study?
A real-world assessment of the Aurora EV-ICD system, including its novel sensing discriminator, is needed to evaluate therapy efficacy and safety through 1 year.
Cohort (n=787)
Yes
In a real-world global registry, the extravascular ICD demonstrated high implant success, effective antitachycardia pacing and shock therapy, and an acceptable safety profile at 1 year.
Captured external expert commentary on this paper, strongest first. Original sources are linked where available.
“The primary results of the pivotal single arm EV-ICD trial reassuringly demonstrated acceptable rates of defibrillation efficacy and freedom from major system or procedure related complications, although inappropriate shocks occurred in 29 of 299 patients (9.7%) and the Kaplan-Meier estimate for inappropriate shocks was 8.5% at 6 months. This elevated rate of inappropriate shocks is similar to early reports of the S-ICD. Interestingly and in contrast with the S-ICD, PWOS was the most common adjudicated mechanism for inappropriate shocks.”
“We are seeing Aurora EV-ICD continue to perform as well in real-world settings as it did in the clinical trial, if not better in some areas. These longer-term outcomes demonstrate consistently strong performance across age groups. Based on the growing body of evidence, the Aurora EV-ICD should be a first-line consideration for patients who haven't had a prior sternotomy and don't require pacing but can benefit from a defibrillator.”
“The strong safety profile and consistent performance of the Aurora EV-ICD in real-world settings reinforces the benefit that patients are experiencing with a defibrillator outside of the vasculature that avoids unnecessary shocks. These real-world outcomes confirm the results observed in the Pivotal Trial showing high defibrillation and ATP success, a low system-related major complication rate, and a reduced rate of inappropriate shocks compared to the Pivotal Trial. We look forward to continued follow up of these patients out to five years.”
Real-world EV-ICD performance data remain limited; leaves open whether premarket safety and efficacy generalize to routine practice.
BACKGROUND: The extravascular implantable cardioverter defibrillator (EV-ICD) was developed to provide defibrillation and antitachycardia pacing therapy without the need for intravascular lead placement. Premarket clinical trials reported EV-ICD safety and efficacy through 3 years, but a real-world assessment of the Aurora EV-ICD system, which includes a novel sensing discriminator, is needed. The objective of this analysis was to assess real-world therapy efficacy and safety of the EV-ICD through 1 year. METHODS: Enlighten: The EV-ICD Post Approval Registry is a prospective, global, real-world registry that included patients who underwent an EV-ICD implantation attempt. The main safety analysis was freedom from all system- or procedure-related major complications through 1 year. Event rates at 1 year were calculated using the Kaplan-Meier method. RESULTS: EV-ICD system implantation was attempted by 168 implanters in 787 patients (49.1±15.2 years old; 28.7% women; 34.1% secondary prevention). Tunneling and lead placement in the substernal space were achieved in 781 of 787 patients (99.2%), with 758 patients (96.3%) remaining implanted following defibrillation and electrical testing. One-year freedom from major system- or procedure-related complications was 89.1% (95% CI, 86.3%–91.3%). The 1-year rate of first appropriate therapy (shock, antitachycardia pacing, or both) was 7.3% (95% CI, 5.5–9.6). Shock success for discrete spontaneous episodes was 100% (61 of 61), and the generalized estimating equations antitachycardia pacing success rate, to account for multiple episodes per patient, was 74.2% for monomorphic ventricular tachycardia. The 1-year rate of first inappropriate shock was 7.1% (95% CI, 5.4–9.4), with noncardiac oversensing as the main cause (40.5% of inappropriately shocked episodes). CONCLUSIONS: In the largest prospective worldwide cohort of patients with an EV-ICD to date, both antitachycardia pacing and shock therapy were effective, while the rate of major complications agreed with the premarket experience through 1 year. These data demonstrate that positive outcomes of the EV-ICD are retained in everyday clinical practice in a large, global cohort with an expanded number of implanters. REGISTRATION: URL: https://www.clinicaltrials.gov ; Unique identifier: NCT06048731.
No takes yet. Share an insight, caveat, or question.
Boersma et al. (2026) conducted a cohort in Indication for implantable cardioverter defibrillator (n=787). Extravascular implantable cardioverter defibrillator (EV-ICD) was evaluated on Freedom from all system- or procedure-related major complications through 1 year (95% CI 86.3-91.3). The extravascular implantable cardioverter defibrillator demonstrated an 89.1% (95% CI, 86.3%-91.3%) freedom from major system- or procedure-related complications at 1 year in a real-world cohort.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: