Answers are grounded in this paper and cite their sources.
A liquid chromatographic mass spectrometric (LC-MS) assay has been developed for cyclosporine A (CyA) in rat plasma using amiodarone as internal standard (IS). Rat plasma (100 microL) containing drug and IS were extracted using liquid-liquid extraction with 4 mL of 95:5 ether:methanol. After evaporation of the organic layer the residue was reconstituted with 500 microL of water. Then the aqueous layer was transferred to LC-MS sample vials. A 10 microL volume was injected. The analysis was performed on a C(8) column 3.5 microm (2.1 x 50 mm) heated to 60 degrees C with a mobile phase consisting of acetonitrile:methanol:0.2% NH(4)OH (60:20:20) at an isocratic flow-rate of 0.2 mL/min. The ions used for quantitation of CyA and IS were m/z 1202.8 and 645.9, with retention times of 3.35 and 4.72 min, respectively. Linear relationships (r(2) > 0.99) were achieved between plasma or blood concentration and peak height ratios (drug:IS) over the concentration range 50-5000 ng/mL. The CV% and mean error were <19%. Based on validation data, the lower limit of quantification for the assay was 50 ng/mL. The reported assay method displayed high measures of linearity, sensitivity, reliability and precision, allowing its applicability in pharmacokinetic studies in rat.
No takes yet. Share an insight, caveat, or question.
Kanduru et al. (2009) studied this question.