Answers are grounded in this paper and cite their sources.
This study used a polyurethane (PU)-based ion selective electrode (ISE) immobilized with di-(2-ethyl hexyl) phosphoric acid (D2EHPA) to detect and quantify Pb 2+ ions at low concentrations (10 −10 to 10 −1 M) in aqueous solution. PU, synthesized from castor oil ( Ricinus communis L.), was utilized as the ISE membrane matrix. The ISE demonstrated a sensitivity of 26.24 ± 0.12 mV/decade, a detection limit of 1.44 × 10 −7 M, and a response time of 10 s. We maintained the measurement stability for up to six days of storage. The results of Pb 2+ quantification produced by ISE were compared with the results using atomic absorption spectroscopy (AAS), indicating similar Pb 2+ concentrations in both artificial and real wastewater samples (calculated t-value < theoretical t-value). Therefore, this ISE is suitable for detecting trace levels of Pb 2+ and quantifying them with high accuracy. • D2EHPA is responsible for attracting the Pb 2+ analyte. • The optimum sensitivity after the addition of TISAB was 27.11 ± 0.11 mV/decade. • The linear range of the D2EHPA/polyurethane-based Pb 2+ ISE was 10 −7 –10 −1 M. • The limit of detection was 1.44 x 10-7 M with a lifetime of 6 days. • Pb 2+ quantification obtained from the ISE was similar to that obtained from AAS.
No takes yet. Share an insight, caveat, or question.
Nisah et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: