Key result
Aortic disc-deployment technique successfully closes post-TAVI VSD without bioprosthetic valve injury.
Why the study?
Ventricular septal defect is a rare complication of transfemoral TAVI with unclear aetiological factors in some cases, requiring novel treatment approaches.
Case Report (n=1)
This case demonstrates a novel technique for closing a post-TAVI VSD by opening the closure device in the ascending aorta to avoid damaging the bioprosthesis.
May warrant vigilance for atypical VSD post-TAVI; leaves open whether the novel closure technique merits prospective evaluation.
Ventricular septal defect (VSD) is a rare complication of transcatheter aortic valve implantation (TAVI) via the transfemoral approach. Aetiological factors leading to VSD have been reported as post-balloon dilatation, oversized prosthesis implantation, and severe calcification of the aorta. However, we present a case of VSD occurring after TAVI with an Edwards Sapien XT prosthesis without any distinct aetiological factors. We used a new technique for closure of the significant VSD; opening the left ventricular disc of the closure device in the ascending aorta and successfully implanting the device without any damage to the bioprosthetic valve.
No takes yet. Share an insight, caveat, or question.
Dursun et al. (2015) conducted a case report in Ventricular septal defect after TAVI (n=1). VSD closure by opening the left ventricular disc of the closure device in the ascending aorta was evaluated on Successful occlusion of VSD without damage to the bioprosthetic valve. A significant ventricular septal defect occurring after TAVI was successfully closed using a new technique of opening the left ventricular disc of the closure device in the ascending aorta.