Answers are grounded in this paper and cite their sources.
The high numbers of mortality related to drug-resistant bacterial superbugs have highlighted the need for rapid, highly sensitive, and cost-effective screening tools, ideally in seamless and point-of-need format. Herein, we report a portable in-flow magnetic transduction-based instrument capable of rapid and highly sensitive detection of whole-bacterial cells in complex matrices. Magnetoresistive (GMR) sensors are combined with microfluidic channels and interfaced with a low noise electronic architecture. The attained practical read-out allows for the interrogation of flowing samples containing small target biological entities (<5 μm). As a proof-of-concept, the device was first optimized for the detection of Klebsiella pneumoniae cells in laboratory samples, demonstrating a 100% sensitivity and 80.8% specificity. Further testing with clinical rectal swabs collected from 40 patients admitted to the hospital emergency department has unveiled its capability for bacteria detection in highly complex matrices with 100% sensitivity, without the need for a pre-enrichment step. Despite some false positives, a screening tool for such application greatly depends on its capacity to correctly identify all positive cases with proper isolation of suspected patients. The test provides a result after 20 min of exposure to 100 μL of sample volume. The magnetic device is a highly flexible tool, being easily customized to detect other bacterial species whenever a corresponding specific recognition ligand is available.
No takes yet. Share an insight, caveat, or question.
Soares et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: