An automated flow injection analysis (FIA) system, based on an initial analyte separation by gas-diffusion and subsequent determination by square-wave voltammetry (SWV) in a flow cell, was developed for the determination of total and free sulfur dioxide (SO(2)) in wine. The proposed method was compared with two iodometric methodologies (the Ripper method and a simplified method commonly used by the wine industry). The developed method displayed good repeatability (RSD lower than 6%) and linearity (between 10 and 250 mg l(-1)) as well as a suitable LOD (3 mg l(-1)) and LOQ (9 mg l(-1)). A major advantage of this system is that SO(2) is directly detected by flow SWV.
No takes yet. Share an insight, caveat, or question.
Gonçalves et al. (2009) studied this question.