This is the authors' abstract. We don't add key points for this paper.
The static and dynamic performances of the RBR argo 3 are investigated using a combination of laboratory-based and in situ datasets from floats deployed as part of an Argo pilot program. Temperature and pressure measurements compare well to co-located reference data acquired from shipboard CTDs. Static accuracy of salinity measurements is significantly improved using 1) a time lag for temperature, 2) a quadratic pressure dependence, and 3) a unit-based calibration for each RBR argo 3 over its full pressure range. Long-term deployments show no significant drift in the RBR argo 3 accuracy. The dynamic response of the RBR argo 3 demonstrates the presence of two different adjustment time scales: a long-term adjustment O (120) s, driven by the temperature difference between the interior of the conductivity cell and the water, and a short-term adjustment O (5–10) s, associated to the initial exchange of heat between the water and the inner ceramic. Corrections for these effects, including dependence on profiling speed, are developed.