Answers are grounded in this paper and cite their sources.
To assess the dosimetric characteristics of HyperArc (HA), noncoplanar VMAT (NC-plan), and coplanar VMAT (HL3arc) in hippocampal-sparing whole-brain radiotherapy (WBRT). Twenty patients undergoing WBRT at our Hospital from June 2021 to March 2023 were selected. The dose parameters of organs at risk (OARs) were evaluated for all three groups, including D 100% , D max , and D mean of hippocampus, D max to the lens, eyes, optic nerves, and chiasm, D mean to the scalp and upper neck, and the dose volume parameters V 5Gy , V 10Gy , V 15Gy , V 20Gy for the upper neck. The D 98% , V 28.5Gy , D 2% , HI, CI, GI for PTV, as well as planning time, and beam-on time. HA had the lowest hippocampal D max and D mean . HA's scalp D mean was lower than HL3arc ( p = 0.007). HA's upper neck region D mean and V 5-20Gy (cm 3 ) were significantly higher than NC-VMAT and HL3arc. For PTV, HA's D 98 and V 28.5Gy were higher than NC-VMAT and HL3arc ( p < 0.05). HA exhibited a significant advantage in terms of HI, CI and GI. The average planning times were 84.67, 17.30, and 20.40 minutes. Among the three treatment approaches, HA demonstrated the lowest bilateral hippocampal doses, but it exhibited the highest average dose and low-dose volume in the upper neck region, with the longest planning optimization process.
No takes yet. Share an insight, caveat, or question.
Yang et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: