Answers are grounded in this paper and cite their sources.
A 6 LiH attenuator of a neutron flux incident on a detector is used to reduce the γ-ray background induced by neutrons in the detector material. This attenuator has been tested during experiments with deuterium (DD) plasmas on the JET tokamak. A specimen of the neutron attenuator with dimensions of ⊘30 × 300 mm has been developed by the Russian Academy of Sciences’ Ioffe Physico-Technical Institute and inserted into a vertical collimator used for γ spectrometry of plasmas. To compare γ-ray spectra recorded with and without the 6 LiH attenuator being mounted, identical discharges with heating of the DD plasma by a neutral particle beam have been selected. For γ rays with energies of <3 MeV, which are induced by neutrons in the detector material, the suppression factor is found to be ∼100. A low attenuation (∼2) observed at energies of >3 MeV can be attributed to the transparency of the 6 LiH attenuator for γ rays. This portion of the spectrum is due to γ radiation of the plasma and γ rays induced by neutrons in the constructional materials of the tokamak. To estimate the efficiency of the 6 LiH attenuators as a mandatory component of the ITER γ-ray diagnostic system, it is necessary that measurements be taken in deuterium-tritium (DT) discharges.
No takes yet. Share an insight, caveat, or question.
Chugunov et al. (2008) studied this question.