Answers are grounded in this paper and cite their sources.
A 27-year-old man with a double inlet left ventricle s/p a classic Fontan palliation and oversewing of a regurgitant tricuspid valve, transposition of the great arteries, and a left persistent superior vena cava (LSVC) underwent pacemaker implantation via the LSVC (Medtronic 4068), with the lead screwed into the coronary sinus ostium in 1997. In 2001, this lead was capped, and a St Jude 1388T lead placed via the LSVC and screwed into the coronary sinus near the first lead (Panel A). In 2015, the patient developed methicillin sensitive Staphylococcal aureus bacteraemia after generator exchange and was referred for system extraction. During laser extraction, lead integrity proximally deteriorated. A Byrd femoral work station was used and advanced to the ostium of the coronary sinus. A Bioptome was used to grasp and extract the most recent lead through the femoral sheath (Panel B). The older lead was not accessible via the work station; therefore, a steerable femoral sheath was used and directed into the LSVC itself. Using the Bioptome and goose neck snare, the lead was successfully extracted (Panel B and see Supplementary material online, Video S1). To our knowledge, this is the first report of a successful femoral extraction of leads placed via an LSVC. Supplementary material is available at Europace online. The full-length version of this report can be viewed at: http://www.escardio.
No takes yet. Share an insight, caveat, or question.
Tanawuttiwat et al. (2016) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: