Key result
Modified-tip fixed-wire PTCA catheters achieve ~85% success for dilating distal coronary lesions.
Why the study?
Lesions in the distal portion of coronary arteries present technological challenges with currently available PTCA equipment, motivating evaluation of new fixed-wire devices with modified distal tips.
Do new fixed-wire PTCA catheters with modified distal tips achieve acceptable procedural success and safety in patients with distal coronary lesions?
Observational (n=323)
Yes
Do new fixed-wire PTCA catheters with modified distal tips achieve acceptable procedural success and safety in patients with distal coronary lesions?
Fixed-wire PTCA catheters with modified distal tips demonstrate an 85% procedural success rate for dilating distal coronary lesions, with an acceptable complication profile.
Answers are grounded in this paper and cite their sources.
May support modified fixed-wire devices for distal lesions; leaves open need for randomized trials versus contemporary equipment.
Lesions in the distal portion of coronary arteries are technological challenges with currently available percutaneous transluminal coronary angioplasty (PTCA) equipment. A study was performed at multiple centers to determine if new “fixed‐wire” devices with a modified distal tip (a 2‐cm flexible tip and 1‐cm standard tip) would be useful in these situations. Three hundred twenty‐three patients were included in the study, 75% male, mean age 63 ± 11 years. Four hundred ninety‐three lesions were attempted; the overall success rate was 85%. Forty‐six (14.2%) clinical complications occurred including: 5 deaths (1.1%); 2 nonfatal myocardial infarctions (0.6%); and 6 emergency coronary artery bypass grafting procedures (2.6%). A device malfunction was noted in 32 patients (9.9%), and included: 8 balloon ruptures or perforations (2.0%); 6 bent or damaged tips (1.9%); and 4 shaft/manifold leaks (1.2%). Two‐month follow‐up assessment for restenosis was done in 352 of 371 (95%) patients and at that time 89% were asymptomatic, had a negative functional test, or patent PTCA site by angiography. The new fixed‐wire PTCA catheters with 1‐ and 2‐cm flexible tips hold promise for dilating lesions in distal locations with an acceptable rate of complications. Additional investigation is needed to determine their role in the treatment of coronary artery disease with PTCA.
No takes yet. Share an insight, caveat, or question.
Talley et al. (1997) conducted an observational in Coronary artery disease with distal lesions (n=323). Fixed-wire PTCA catheters with modified distal tips was evaluated on Overall success rate. Fixed-wire PTCA catheters with modified distal tips achieved an 85% overall success rate for dilating distal coronary lesions, with a 14.2% clinical complication rate.
Synapse has enriched 10 related papers. Consider them for comparative context: