A French national quality control study for the serological and molecular diagnosis of hepatitis delta virus (HDV) was organized. Total HDV antibodies were properly detected by all laboratories; 8/14 laboratories failed to detect low titers of IgM, and 6/11 failed to quantify and/or underestimated the RNA viral load in several samples. These discrepancies are likely related to the molecular diversity of HDV.