Key result
Memory loop-recorders are associated with double the diagnostic ECG accuracy of wrist-worn devices.
Why the study?
The study aimed to evaluate the effectiveness of a patient-actuated transtelephonic cardiac monitoring system for documenting arrhythmias and pacemaker function.
Do transtelephonic monitoring devices accurately document cardiac arrhythmias and pacemaker function in patients with pacemakers or symptomatic arrhythmias?
Observational (n=82)
Do transtelephonic monitoring devices accurately document cardiac arrhythmias and pacemaker function in patients with pacemakers or symptomatic arrhythmias?
A memory loop-recording transtelephonic monitoring system provides highly accurate ECG diagnosis for patients with pacemakers or symptomatic arrhythmias, whereas a wrist-worn recorder demonstrated low accuracy.
Background: The purpose of this trial was to evaluate a patient actuated transtelephonic cardiac monitoring system in order to document cardiac arrhythmias and cardiac pacemaker function. Material: Eighty‐two patients were prospectively evaluated, 69 with pacemaker (group I) and 13 with symptomatic arrhythmias (group II). Two different recorders were used: a memory loop‐recorder (KH) and a wrist‐worn recorder (HW). Both of them were implemented using a small, portable, battery‐powered transmitter, which monitors a modified V5 in KH and a lead 1 with the HW over regular nondigital pulse telephone lines. Results: In group I, 54 patients used single chamber pacemakers, 2 VDDR, and 13 used dual chamber devices. In group II, all patients included referred palpitations as their symptom. In group I, 248 registers were made using the KH and 50 with the HW. Recordings were made with KH had a 96% accuracy in the diagnosis of the ECG, while HW recordings failed to detect the QRS and the spike in 52% of the cases. Arrhythmia patients (group II) made 65 recordings with KH, all symptomatic: 28 were ventricular ectopic beats (23 isolated, 4 bigeminy, and 1 coupled ventricular ectopic beats); 9 supraventricular ectopic beats (isolated); 18 episodes of sinus tachycardia; and 10 normal sinus rhythm. Artifact was present partially in 6 other recordings, but did not affect the diagnosis. Two patients made no recordings and were excluded from the trial (group I). Conclusions: The loop‐recording transtelephonic monitoring system is an excellent tool for the evaluation of patients with symptomatic arrhythmias and pacemaker. The memory KH had an excellent performance, even over regular telephone lines. The accuracy of the HW recordings was low and failed to evaluate the QRS, probably due to the direction of the AQRS vector.
No takes yet. Share an insight, caveat, or question.
Dubner et al. (1996) conducted an observational in Pacemaker or symptomatic arrhythmias (n=82). Memory loop-recorder (KH) vs. Wrist-worn recorder (HW) was evaluated on Accuracy in the diagnosis of the ECG. The memory loop-recorder (KH) achieved 96% accuracy in ECG diagnosis, whereas the wrist-worn recorder (HW) failed to detect the QRS and pacemaker spike in 52% of cases.