Key result
Pacemaker telemetry using STOP-AF criteria achieved only 57% sensitivity and 64% specificity for detecting atrial fibrillation compared to 24-hour Holter recordings.
Why the study?
Does pacemaker telemetry accurately detect atrial fibrillation compared to 24 h Holter recordings in patients with permanent pacemakers?
Observational (n=18)
Does pacemaker telemetry accurately detect atrial fibrillation compared to 24 h Holter recordings in patients with permanent pacemakers?
Pacemaker telemetry using STOP-AF criteria demonstrated low sensitivity and specificity for detecting atrial fibrillation compared to Holter monitoring, indicating a need for further validation of device-specific diagnostic criteria.
Answers are grounded in this paper and cite their sources.
Pacemaker telemetry may be unreliable for AF detection; leaves open need for validated device-specific criteria versus ambulatory monitoring.
AIMS: To compare pacemaker telemetry with simultaneous Holter recordings in the diagnosis of atrial fibrillation and to evaluate the STOP-AF study telemetry criteria for the presence of atrial fibrillation. METHODS AND RESULTS: 18 consecutive patients enrolled in the STOP-AF study had simultaneous 24 h Holter recordings and down-loaded pacemaker telemetry. There was good agreement on heart rate, but the STOP-AF pacemaker criteria achieved only 57% sensitivity with 64% specificity for the presence of atrial fibrillation on Holter recording over 24 h. False-positives appeared to result from far-field sensing while false-negatives occurred with very short episodes of atrial fibrillation. The pacemaker's antitachycardia responses were not specific for the presence of atrial fibrillation. CONCLUSION: Pacemaker telemetry is a potentially important source of data on cardiac arrhythmias. Further studies are required to define the limitations of these data in specific devices before they can be interpreted with confidence.
No takes yet. Share an insight, caveat, or question.
Chris Plummer (2001) conducted an observational in Atrial fibrillation (n=18). Pacemaker telemetry vs. Holter recordings was evaluated on Diagnosis of atrial fibrillation. Pacemaker telemetry using STOP-AF criteria achieved only 57% sensitivity and 64% specificity for detecting atrial fibrillation compared to 24-hour Holter recordings.