Key result
Novel ultrasonic-Doppler method accurately detects subtle aortic wall vibrations despite large heartbeat motion.
Why the study?
Accurate detection of small vibrations of the aortic wall near the heart is necessary for acoustical diagnosis of atherosclerosis but is difficult with previous ultrasonic methods due to large heartbeat motion.
A new ultrasonic-Doppler-based method was developed and evaluated in simulations to accurately measure small vibrations of the aortic wall for diagnosing atherosclerosis.
Answers are grounded in this paper and cite their sources.
Simulation-validated method may enable aortic vibration assessment for atherosclerosis; leaves open in vivo and clinical validation.
To diagnose atherosclerosis based on the acoustic characteristics of the aortic wall, it is necessary to detect vibration signals from various parts of the aortic wall near the heart. It is, however, difficult to obtain such small vibrations using previously proposed ultrasonic diagnostic methods or systems because the small vibrations are superimposed on the motion with large amplitude due to the heartbeat. The authors developed a new method for overcoming this problem and accurately measuring small vibrations of the aortic wall using ultrasound. The Letter describes the accuracy of evaluation using this new method by simulation experiments using a system which simulates small vibrations of the aortic wall near the heart.
No takes yet. Share an insight, caveat, or question.
Kawabe et al. (1993) conducted a letter in Atherosclerosis. New ultrasonic-Doppler-based method for measuring small vibrations was evaluated on Accuracy of measuring small vibrations of the aortic wall. A new ultrasonic-Doppler-based method was developed and evaluated in simulation experiments to accurately measure small vibrations of the aortic wall despite large amplitude motion from the heartbeat.
Synapse has enriched 10 related papers. Consider them for comparative context: