Key result
Arterial pressure-guided VV optimization outperforms ECG and LVOT VTI methods, achieving 85% optimum singularity.
Why the study?
The internal validity of echocardiographic, electrocardiographic, and haemodynamic methods for VV optimization in AF patients was uncertain.
Does non-invasive arterial pressure provide more reliable VV optimization compared to LVOT VTI and ECG in patients with atrial fibrillation and biventricular pacemakers?
Observational (n=20)
Does non-invasive arterial pressure provide more reliable VV optimization compared to LVOT VTI and ECG in patients with atrial fibrillation and biventricular pacemakers?
Non-invasive arterial pressure provides a more reliable and valid method for VV optimization in biventricular pacing compared to echocardiographic LVOT VTI or ECG QRS minimization.
Answers are grounded in this paper and cite their sources.
VV optimization methods lack consistent internal validity in AF; leaves open optimal approach for biventricular pacing.
BACKGROUND: In atrial fibrillation (AF), VV optimization of biventricular pacemakers can be examined in isolation. We used this approach to evaluate internal validity of three VV optimization methods by three criteria. METHODS AND RESULTS: Twenty patients (16 men, age 75 ± 7) in AF were optimized, at two paced heart rates, by LVOT VTI (flow), non-invasive arterial pressure, and ECG (minimizing QRS duration). Each optimization method was evaluated for: singularity (unique peak of function), reproducibility of optimum, and biological plausibility of the distribution of optima. The reproducibility (standard deviation of the difference, SDD) of the optimal VV delay was 10 ms for pressure, versus 8 ms (p=ns) for QRS and 34 ms (p<0.01) for flow. Singularity of optimum was 85% for pressure, 63% for ECG and 45% for flow (Chi(2)=10.9, p<0.005). The distribution of pressure optima was biologically plausible, with 80% LV pre-excited (p=0.007). The distributions of ECG (55% LV pre-excitation) and flow (45% LV pre-excitation) optima were no different to random (p=ns). The pressure-derived optimal VV delay is unaffected by the paced rate: SDD between slow and fast heart rate is 9 ms, no different from the reproducibility SDD at both heart rates. CONCLUSIONS: Using non-invasive arterial pressure, VV delay optimization by parabolic fitting is achievable with good precision, satisfying all 3 criteria of internal validity. VV optimum is unaffected by heart rate. Neither QRS minimization nor LVOT VTI satisfy all validity criteria, and therefore seem weaker candidate modalities for VV optimization. AF, unlinking interventricular from atrioventricular delay, uniquely exposes resynchronization concepts to experimental scrutiny.
No takes yet. Share an insight, caveat, or question.
Kyriacou et al. (2012) conducted an observational in Atrial fibrillation (n=20). Non-invasive arterial pressure for VV optimization vs. LVOT VTI (flow) and ECG (minimizing QRS duration) was evaluated on Singularity of optimum (p=<0.005). VV delay optimization using non-invasive arterial pressure achieved 85% singularity of optimum, outperforming ECG (63%) and LVOT VTI (45%) methods (P<0.005).