Key result
Noninvasive optimization of atrioventricular and interventricular delays using impedance cardiography significantly increased cardiac output by 32.6% compared to intrinsic rhythm.
Why the study?
Does individual noninvasive optimization of AV and VV delays using impedance cardiography improve cardiac output in patients with advanced heart failure after CRT implantation?
Observational (n=20)
No
Does individual noninvasive optimization of AV and VV delays using impedance cardiography improve cardiac output in patients with advanced heart failure after CRT implantation?
Answers are grounded in this paper and cite their sources.
Individual optimization of AV and VV delays using noninvasive impedance cardiography significantly improves cardiac output compared to intrinsic rhythm and factory settings in CRT patients.
BACKGROUND: Device optimization is not routinely performed in patients who underwent cardiac resynchronization therapy (CRT) device implantation. Noninvasive optimization of CRT devices by measurement of cardiac output (CO) can be used as a simple method to assess ventricular systolic performance. The aim of this study was to assess whether optimization of atrioventricular (AV) and interventricular (VV) delay can improve hemodynamic response to CRT and whether this optimization should be performed for each patient individually. METHODS: Twenty patients with advanced heart failure New York Heart Association (NYHA) class III/IV, left ventricular ejection fraction ≤ 35% and left bundle branch block (QRS ≥ 120 ms) in sinus rhythm were evaluated from 24 h to 48 h after implantation of a CRT device by means of impedance cardiography (ICG). CO was first measured at each patient's intrinsic rhythm. Patients then underwent adjustments of AV and VV delay from 80 ms to 140 ms and from -60 ms to +60 ms, respectively in 20 ms increment steps and CO at each setting was measured by ICG. Both AV and VV delays were programmed according to the greatest improvement in CO compared to intrinsic rhythm. RESULTS: There was a statistically signifi cant increase in CO measured at the intrinsic rhythm compared to different AV delay by mean of 21% (3.8 ± 1.0 vs. 4.6 ± 0.1 L/min, p < 0.05). Optimal AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing caused additional increased CO from intrinsic rhythm by mean of 32.6% (3.8 ± 1.0 vs. 5.04 ± ± 1.0 L/min, p < 0.05). Optimal AV/VV setting delays also resulted in improved hemodynamic responses compared to VV factory setting delay. CONCLUSIONS: Both AV and VV delay optimization should be performed in clinical practice. Optimal AV delay improved outcome. However, combination of optimized AV/VV delays provided the best hemodynamic response. Optimized AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing increased hemodynamic output compared to intrinsic rhythm and VV factory setting delay.
No takes yet. Share an insight, caveat, or question.
Urbanek et al. (2013) conducted an observational in Advanced heart failure (n=20). Noninvasive optimization of AV and VV delays via impedance cardiography vs. Intrinsic rhythm was evaluated on Cardiac output (L/min) (32.6% increase, p=<0.05). Noninvasive optimization of atrioventricular and interventricular delays using impedance cardiography significantly increased cardiac output by 32.6% compared to intrinsic rhythm.