Answers are grounded in this paper and cite their sources.
Inorganic CsPbBr3 perovskite devices have been manufactured and tested as dosimeters under both conventional and Intensity Modulated Radiotherapy (IMRT) X-ray beams. Samples showed a very good linear dependence of the collected charge/current on dose/dose rates in the range of 0.1–5.0 Gy/0.1–4.0 Gy/min of interest for clinical applications. A device sensitivity of about 70 nC/Gy mm3 compares favorably with other solid-state dosimeters. The first verification of an IMRT dose profile of a prostate cancer treatment, performed by moving the perovskite device on a 10 cm-long profile with a 0.5 mm pitch, showed agreement within 5% with the dose distribution required by the treatment planning system.
No takes yet. Share an insight, caveat, or question.
Bruzzi et al. (2019) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: