Key result
Reduced frame rate protocol during CIED implantation cuts radiation exposure ~62% without increasing complications.
Why the study?
Radiation exposure during cardiac implantable electrical device implantation remains a substantial health hazard, and low-cost dose reduction methods are needed.
Does a reduced frame rate protocol (3.8 frames/s) reduce radiation exposure and maintain safety compared to standard frame rate (7.5 frames/s) in patients undergoing CIED implantation?
Cohort (n=495)
No
Does a reduced frame rate protocol (3.8 frames/s) reduce radiation exposure and maintain safety compared to standard frame rate (7.5 frames/s) in patients undergoing CIED implantation?
Halving the fluoroscopy frame rate to 3.8 frames/s during CIED implantation significantly reduces radiation exposure by approximately 62% without increasing procedural complications.
Reduced frame rate may lower radiation exposure in CIED implantation; leaves open effects on complications and procedural success.
Background Radiation exposure to patient and surgeon during cardiac implantable electrical device (CIED) procedures remains a substantial health hazard to date. Advanced technical options for radiation dose reduction often pose considerable financial hurdles. We propose a near-zero cost, low-effort modification to a clinical x-ray system significantly reducing radiation dose during CIED implantation. Objective We aim to evaluate a reduced frame rate protocol in CIED implantation for complication rates and reduction in radiation exposure. Methods Starting May 2019, the frame rate during CIED implantations at our hospital was halved from 7.5 frames/s to 3.8 frames/s, and no further technical changes were made. During the following year, 264 patients were operated using this protocol and retrospectively compared with 231 cases implanted in the year before the protocol change, totaling 495 cases. Of these, 17%, 63%, and 19% were single-chamber, dual-chamber, or resynchronization devices, respectively. Incidence of complication prior to hospital discharge was considered the primary endpoint of the analysis. Radiation dose and procedural parameters were secondary endpoints. Results There was no increase in complications with the reduced frame rate protocol. Regression analysis further supported that the reduced frame rate radiation protocol was not associated with complication rates. Radiation exposure measured as dose area product was significantly reduced by ∼62% (median 369 [interquartile range 154–1207] cGy·cm 2 via the reduced frame rate protocol vs median 970 [interquartile range 400–1906] cGy·cm 2 with the standard frame rate; P < 0.01). Conclusion A reduction of frame rate during CIED implantation is safe in terms of complication incidence and effective in terms of reducing radiation exposure.
No takes yet. Share an insight, caveat, or question.
Bork et al. (2023) conducted a cohort in Cardiac implantable electrical device (CIED) implantation (n=495). Reduced frame rate protocol (3.8 frames/s) vs. Standard frame rate (7.5 frames/s) was evaluated on Incidence of complication prior to hospital discharge. A reduced frame rate protocol (3.8 frames/s) during CIED implantation did not increase complications and significantly reduced radiation exposure by ~62% (median 369 vs 970 cGy·cm2; P<0.01).